DltDlyD
) using the MeterLink Transducer Swap-out Wizard (see
in Determining
L Value).
Average delay time and delta delay time modifications
The transducer pair average delay time and delta delay time are located on the transducer
pair calibration sheet. These values must be downloaded to the appropriate meter data
points (
AvgDlyA... AvgDlyD, DltDlyA... DltDlyD
). The lengths of the transducers are also
included on the calibration sheet and are etched on the transducers. Likewise the lengths
of the stalk assemblies, transducer holders, and mounts are etched on the individual
components. The length of the meter body is found on the original calibration sheet
supplied with the meter.
Determining the “L” value
The value "L" is determined by adding the length of the meter body to the lengths of the
two mounts and subtracting the lengths of the transducer holders, stalk assemblies, and
transducers. This value should be written to the appropriate meter data points for each
chord that received new transducers (
LA... LD
for the “L” dimension
calculation.
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
52
Emerson.com/Rosemount