Table 2-3: Meter Monitor maintenance
(continued)
MeterLink utility
Diagnostics
Action(s)
•
Collect a complete Archive
log and contact your local
area Emerson Flow service
representative.
MeterLink Meter Monitor
(Summary) view
•
Chord failure MeterLink
•
The meter is unable to
obtain measurement data
from a pair of transducers.
•
The cause may be isolated
to one pair of transducers
or its cabling. Check the
transducer wiring for this
pair of transducers to make
sure connections are secure
and wired correctly.
•
Verify the average gain of
this transducer pair is not
above 90dB. Read the value
from the MeterLink Monitor
Page or using AMS under
Service Tools
→
Path
performance.
•
Remove the transducer and
clean the transducer face
(Transducer removal
procedure).
Security seals
•
Endcap seals
•
Endcaps latches
•
Transmitter Electronics
Enclosure
•
Base Enclosure
•
Shroud seals
•
Only authorized personnel
may remove security seals.
Follow your standard
operating procedure to
report seals that have been
tampered with or removed
and replace the seals per
instructions in "Security seal
installation" in the
Installation manual
(00825-0600-3104).
External ground wiring
•
Transmitter Electronics
Enclosure ground lug
•
Inspect ground lug wiring
and make sure the wiring is
tightly secured.
Troubleshooting
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104