Table 2-3: Meter Monitor maintenance
(continued)
MeterLink utility
Diagnostics
Action(s)
MeterLink Logs/ Reports Menu
•
Meter performed a Warm
start or a Warm start
required
•
Meter performed a Warm
start:
—
Collect an Archive event
log (Auditlog) using
MeterLink to view
configuration
parameter changes and
when they changed.
•
Warm start is required:
—
When you make
changes to the
transducer
characteristics, sample
rates, the device
number, or a Modbus
map file.
MeterLink
Tools
→
Edit/
Compare Configuration
menu
•
Meter performed a Cold
start
•
The meter configuration
has reset to default values
and the meter is not
configured correctly to
measure flow.
•
Unless the cold start
occurred after upgrading
firmware, you may need to
replace the CPU Module.
•
If the Cold Start occurred
after a firmware upgrade,
you must reconfigure the
meter from a previously
saved configuration file
using the
Edit
→
Compare
Configuration
screen. Then
clear the latched alarm on
the Status Summary page.
MeterLink Logs/ Reports Menu
•
Power failure
•
If this was a known power
fail or restart of the meter
just acknowledge this alarm
on the Status Summary
page.
•
If this was an unexpected
restart of the meter, verify
the integrity of the power
to the meter and make sure
that the voltage level is in
the range of 11-36 VDC at
the meter.