Table 2-3: Meter Monitor maintenance
(continued)
MeterLink utility
Diagnostics
Action(s)
Meter Monitor (Detailed) view
Meter Data List
Electronics Temperature of out
range
•
Temperature of the
electronics is out of
nominal operating range
below -40 °C or above 100
°C (-40 °F or above 212 °F).
Meter Monitor (Detailed) view
Meter Data List
Electronics voltage out of range
•
Systems voltages are valid if
1.0V, 1.2V, 2.5V, 3.3V or
the Acquisition Module
valid voltages are 1.2V,
2.5V or 3.3V.
•
Replace the CPU Module if
one or more of the System
Voltages is out of range.
•
Replace the Acquisition
Module if one or more of
the voltages is out of range.
MeterLink Tools Menu
•
Frequency output
•
Run the Frequency Outputs
test
•
If the output reads zero,
you may require a pull up
resistor 1.2k OHM, 0.5W.
Maintenance and Troubleshooting Manual
Troubleshooting
00809-0600-3104
April 2022
Rosemount Gas Ultrasonic Flow Meters