drained and/ or pressure has not been relieved from the meter. Correct the issue
before attempting to remove the holder.
6. The transducer holder has now been removed from the meter with the transducer
still installed inside the transducer holder.
7. Make a note of the removed transducer holder length which is used to update the
meter configuration during the Transducer Swap-out procedure in MeterLink, after
all of the transducer holders are replaced.
Figure 3-7: Transducer holder length and set screw identification
A. Transducer holder set screws
B. Transducer holder length identification
8. Loosen the three Allen setscrews with a 1/16” hex driver securing the transducer
assembly and stalk, if installed. Carefully remove the transducer by pulling it from
the T-Slot transducer holder (or stalk if installed) without rotating.
9. Clean the holder with a dry cloth.
3.3.3
Install the transducer holder with the extractor tool
To replace transducer holder while under pressure, refer to Ultrasonic Split Clamp
Extractor Tool Operation Manual (00809-0200-3417).
Note
The extractor tool is used on Rosemount
™
3414 Ultrasonic Flow Meters and is not
applicable for use with Rosemount 3412 and 3411 Ultrasonic Flow Meters.
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104