Table 2-2: Troubleshooting
(continued)
Error
Recommended action(s)
Waveform contains an
excessive amount of
noise
•
Use the MeterLink
Meter
→
Signal Analyzer
to increase the
StackSize
until noise level decreases (settings can be 1 (none) 2, 4,
8, or 16). If increasing the
StackSize
is not successful, try turning on
the filter or consult with Emerson Customer Support if you are
unsure of how stacking a signal can affect the meter's operation.
2.2.1
Meter monitoring maintenance
The Monitor (Summary) includes the direction of flow measurement, velocity rate, units of
measurement, uncorrected or corrected flow (if applicable for your meter) and a bar graph
for a visual comparison between the velocities for each chord. This is the default view
displayed when you select
Meter
→
Monitor
from the toolbar.
Figure 2-4: Meter Monitor (Summary) view
Run MeterLink and open the Meter Monitor (Detailed) view to perform a diagnostics
health check and or adjust parameters for your site requirements. If you wish to use the
Monitor (Detailed) dialog as the default view, click the checkbox in the lower portion of
the dialog box.
Troubleshooting
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
26