Figure 3-5: T-22 transducer assembly, holder, transformer assembly, retainer,
cable nut and chordset
A. T-21 transformer assy. (W-01 P/N 1-360-03-090) or T-22 transformer assy. (W-02
P/N 1-360-03-110)
B. Transducer holder (Type - H1 P/N 1-360-01-128, H2 P/N 1-360-01-228)
C. Mount (Inconel mount and holder)
D. Transducer port (meter body)
E. Transducer retainer (Standard P/N 1-360-01-958) (High temperature P/N
1-360-01-978)
F. Cable nut and chordset:
•
100° C -5’ length (P/N 3-3400-190)
•
100° C -15’ length (P/N 3-3400-194)
9. For T-21 and T-22 transducers, place the retainer over the transformer assembly.
Ensure the retainer threads are aligned correctly and hand-tighten. Use a 1
⅛
”
wrench and turn clockwise until the transformer assembly is fully seated in the
transducer holder.
Note
Do not over tighten the retainer.
10. Reconnect the transducer chordset to the transducer holder or the retainer if
installing T-21 or T-22 transducers. The internal connector of the transducer
chordset is keyed and will only go on one way. Secure the transducer cabling nut by
turning clock-wise. Ensure the cable nut threads are correctly aligned.
11. Repeat
for the remaining transducer assemblies which were
replaced.