Update the calibration parameters for every chord when replacing a pair of
transducers. Refer to
Modifying the calibration parameters for T-Slot transducers
.
Table 3-2: Tools and supplies for disassembly and assembly
Tool list, shop supplies and reference documents
3/8” drive ratchet
3/8” drive extension
3/8” drive 7/16” deep socket
3/8” drive 5/16” hex socket
3/8” drive 1/2” hex socket
3/8” to 1/2” driver adapter
1/2” drive ratchet
1/2” drive 1
¼
” Socket
1
¼
” combination wrench
1
⅛
” combination wrench
1” combination wrench
3/16” flat Blade screw driver
1/16” Xcelite Hex driver
6 mm Allen wrench
N.A.S. Loctite Nickel Anti-Seize 16 oz. Brush Top P/N 77164
Molykote 111
Shop supplies
Shop rags
Adhesive stickers for the cable nuts Sharpie Fine Point Permanent Marker (Black) Series # 30000, to
label transducers:
•
A1, A2, B1, B2, C1, C2, D1 and D2 for SeniorSonic meters
•
A1, A2, B1 and B2 for JuniorSonic meters
Liquid leak detection fluid (customer supplied item)
Procedure
1. Blow the line down according to the site standard operating procedures.
2. Ensure that the line pressure is down to atmospheric pressure prior to disassembly.
3. Disconnect transducer cabling and chordset from the transducer assembly by
turning the cable nut counter-clock wise.
4. Disconnect the transducer cable nut. Pull the chordset from the transducer
assembly.
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
46