Chart -
the chart utility displays the data collected for the value selected from the Chart
drop-down list.
Refer to
for meter maintenance hardware
diagnostics.
Table 2-3: Meter Monitor maintenance
MeterLink utility
Diagnostics
Action(s)
Meter Monitor (Summary) view Check Status for active alarms
•
Meter Status LED is green if
there are no active alarms.
This indicates the meter is
measuring flow and
operating within the
calibrated parameters.
•
Meter Status LED is red.
This indicates an active
alarm. Resolve and
acknowledge active alarms
as displayed on the Status
Summary page. Click the
Help
button, , beside the
alarm description to display
information about the
alarm and recommended
actions to resolve the issue.
Troubleshooting
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
28