2.2
Troubleshooting the meter
and the following sections show errors that may occur with the meter hardware,
firmware or connections and recommend actions to resolve the problem(s).
Table 2-2: Troubleshooting
Error
Recommended action(s)
Acquisition Module Error
•
Check/Replace I.S. Barrier.
•
Check/Replace Acquisition module.
•
Attempt the Program Download procedure to install the firmware.
—
Cycle power to the meter
—
Replace the Acquisition Module.
—
If the Acquisition Module cannot be reprogrammed, collect a
complete Archive log and contact your local area Emerson Flow
service representative.
Acquisition Module is not
compatible with
firmware
•
Check Acquisition Module is compatible with Model # configured in
meter (i.e. 3414 4-path or 3418 8-path).
•
Replace the Acquisition Module.