Instruction Manual
IM-106-350, Rev 2.2
July 2008
2-9
Oxymitter 5000
Figure 2-7. Orienting the
Optional Vee Deflector
7. Insert probe through the opening in the mounting flange and bolt the unit
to the flange. When probe lengths selected are 9 to 18 ft (2.74 to
5.49 m), special brackets are supplied to provide additional support for
the probe inside the flue or stack (Figure 2-6).
NOTE
If process temperatures will exceed 392°F (200°C), use anti-seize compound
on stud threads to ease future removal of Oxymitter 5000. For ambient
temperatures that will exceed 185°F (85°C), we reco mmend the remote
mounted electronics option.
8. If insulation is being removed to access the duct work for Oxymitter
5000 mounting, make sure the insulation is replaced afterward
(Figure 2-9).
Vee
Deflector
Vee
Deflector
Diffusion
Element
Setscrew
Filter
Gas Flow
Direction
Apex
38730003
Uninsulated stacks or ducts may cause ambient temperatures around the electronics to
exceed 185°F (85°C), which may cause overheating da mage to the electronics.
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 12: ......
Page 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...