Instruction Manual
IM-106-350, Rev 2.2
July 2008
E-17
Oxymitter 5000
Mode sheds
from RCAS to
AUTO
1. Remote Cascade Value
2. Shed Timer
1. Host system is not writing RCAS_IN with a
quality and status of “good cascade” within shed
time (see 2 below).
2. The mode shed timer, SHED_RCAS in the
resource block is set too low. Increase the value.
Mode sheds
from ROUT to
MAN
1. Remote output value
2. Shed timer
1. Host system is not writing ROUT_IN with a
quality and status of “good cascade” within shed
time (see 2 below).
2. The mode shed timer, SHED_RCAS, in the
resource block is set too low. Increase the value.
Process and/or
block alarms
will not work
1. Features
2. Notification
3. Status Options
1. FEATURES_SEL does not have Alerts
enabled. Enable the Alerts bit.
2. LIM_NOTIFY is not high enough. Set equal to
MAX_NOTIFY.
3. STATUS_OPTS has Propagate Fault Forward
bit set. This should be cleared to cause an alarm
to occur.
Symptom
Possible Causes
Corrective Action
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 12: ......
Page 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...