8-14
Instruction Manual
IM-106-350, Rev 2.2
July 2008
Oxymitter 5000
Figure 8-10. Fault 8, Low Heater Temp
Fault 8, Low Heater Temp
Figure 8-10 shows the electronic assembly for
an Oxymitter 5000 with a membrane keypad
(upper view) and an Oxymitter 5000 with an LOI
(lower view).
Membrane Keypad
When Fault 8 is detected, the HEATER LED
flashes four times, pauses for three seconds,
and repeats.
1. The low heater temperature alarm is active
when the thermocouple reading has
dropped below 28.6 mV.
2. If the thermocouple reading continues to
ramp downward for one minute and does
not return to the temperature set point of
approximately 29.3 mV, then an Open
Heater fault will be displayed.
3. Power down the electronics. Remove the
electronic assembly per "Electronic
Assembly Replacement" in Section 9:
Maintenance and Service. Using a
multimeter, measure the resistance across
the terminals of heater connector, J8.
4. If the heater is good, the reading will be
approximately 70 ohms. If the heater is
open, see "Heater Strut Replacement" in
Section 9: Maintenance and Service.
LOI
When Fault 8 is detected, the LOI displays the
"O2 Temp Low" message. Refer to the
comments and procedures in paragraphs 1
through 4 above.
38730079
LOI
KEYPAD
Alarms
O2 Temp Low
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATER T/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATER T/C +
HEATER T/C -
INC
INC
DEC
DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
SW2
TP1
J1
TP2
TP3
RED
YEL
GRN
ORG
TP4
TP5
TP6
ON
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 12: ......
Page 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...