Oxymitter 5000
8-2
Instruction Manual
IM-106-350, Rev 2.2
July 2008
Figure 8-1. O
2
Sensor mV
Reading vs. % O
2
at 1357°F
(736°C) (Reference Air, 20.9%
O
2
)
38730096
100
10
1
0.01
0.1
0
0
50
100
150
200
Concentration O (%)
2
EMF(mV)
O Sensor Performance at 1357
2
O
F (736
O
C)
O
2
%
100
20
15
10
9
8
7
6
5
4
EMF(mV)
-34
1.0
7.25
16.1
18.4
21.1
23.8
27.2
31.2
36.0
O
2
%
3
2
1
0.8
0.6
0.5
0.4
0.2
0.1
0.01
EMF(mV)
42.3
51.1
66.1
71.0
77.5
81.5
86.3
101.4
116.6
166.8
Summary of Contents for Oxymitter 5000
Page 2: ......
Page 6: ......
Page 12: ......
Page 22: ...Oxymitter 5000 xii Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 42: ...Oxymitter 5000 1 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 62: ...Oxymitter 5000 2 20 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 74: ...Oxymitter 5000 4 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 78: ...Oxymitter 5000 5 4 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 94: ...Oxymitter 5000 7 6 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 140: ...Oxymitter 5000 9 22 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 184: ...Oxymitter 5000 B 2 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 204: ...Oxymitter 5000 D 14 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 222: ...Oxymitter 5000 E 18 Instruction Manual IM 106 350 Rev 2 2 July 2008 ...
Page 224: ...Instruction Manual IM 106 350 Rev 2 2 July 2008 Index 2 Oxymitter 5000 ...