RSTi-OM User Manual
Section 8
GFK-3212A
Jun 2021
Utilizing IOLM Features
81
Although you can make configuration changes using the web page, PROFINET IO configuration
parameters overwrite the values on the
IO-Link Settings
page. See
IO-Link Port Settings (IO-Link Port
for PROFINET IO configuration procedures.
Use this procedure to configure device validation.
1.
Click Configuration | IO-Link Settings.
2.
Click the EDIT button.
3.
Select Compatible or Identical for the Device Validation mode.
Note: Identical Device Validation requires a device serial number to operate.
4.
Click the GET ATTACHED button or manually complete the Vendor ID, Device, ID, and serial
number.
5.
If the device does not have a serial number, you should not select Identical because the IOLM
requires a serial number to identify a specific device.
Figure 68: GET ATTACHED Button
6.
Click the SAVE button. If the wrong or incompatible device is connected to the port, the IO-Link
port LED flashes red and no IO-Link activity occurs on the port until the issue is resolved.