2009 by ELTRA GmbH Germany – July 2009 – Service Manual Thermostep
Page 45
Fig. 3.8.5: LHS – Disconnecting purge tube of the balance
g.)
From the LHS undo the wing nut which secures the wind shield glass tube. Allow
glass tube to slide down and rest on the balance shield (see Fig. 1.5).
h.)
Using a small 5.5mm nut driver loosen the four 5.5mm nuts holding the clamps
that secure the balance shield to the balance. Turn each clamp 90 degrees and from
the LHS carefully lift off the balance shield and wind shield.
i.)
Remove completely 2 x 4mm allen screws and 2 x aluminum disc from the front
and rear of the balance base plate.
Fig. 3.8.6: LHS/RHS – Loosen the balance shield