2009 by ELTRA GmbH Germany – July 2009 – Service Manual Thermostep
Page 42
Fig. 3.7.2: FS – Hinge out front metal plate
c.)
Unfasten the screw that fixes the cylinder at the instrument and remove the cylinder.
Fig. 3.7.3: FS – Screw of cylinder of internal blower
d.)
Insert the new cylinder and fix it.
e.)
Close the front metal plate again.
3.8
Change of balance
Required tools:
4 mm Allen Key
5.5 mm Nut Driver (short)
15 cm length of Heavy Duty Cardboard Tube (for lid support)
Small Phillips Screwdriver (for data plug)
Unplug instrument from electrical supply and disconnect the compressed air supply.
a.)
Prop up lid using a section of heavy duty cardboard tube under the LHS pneumatic
ram arm.