2009 by ELTRA GmbH Germany – July 2009 – Service Manual Thermostep
Page 22
Check if setting of serial port (COM2) at TGA software is correct (see Operation manual:
Form –
Login
Management
).
Check if setting of serial port, where the balance cable is plugged in, at PC hardware is
correct.
Check if the balance is switched on! The access to balance display is at the RHS.
Fig. 2.4: RHS – Balancedisplay and -operationelements
Check if balance is programmed correctly (see
1.6
)!
At c.)
Error message: “Balance doesn't report stable weight!”
Check if balance pedestal is fitted at the extension correctly! Check if extension is mounted
inside the analyser correctly.
Remove possible disturbing influences at the balance pedestal!
Tare the balance at the sorftware (see Operation manual: Form –
Balance
). If this is not
possible tare the balance manually via pressing the tare button beside the balance display. The
access to balance display is at the RHS. Taring of balance is indicated by an accustic signal
and 0.0000 at the display.
Zu d.)
Error message: “No communications! Please check cable, port settings and instrument.”