2009 by ELTRA GmbH Germany – July 2009 – Service Manual Thermostep
Page 50
Fig. 3.9.6:: RHS – Removing of the Thermocouple out of the ceramic tube
h.)
Insert the cable of the new thermocouple through the rubber grommets along the frame
and fix it with the fitting.
i.)
Fasten the wires at the socket and remind their orientation. Plug in the socket.
j.)
Push the thermocouple carefully till the bottomed in the ceramic device.
k.)
Close the upper right isolation metal plate and side panel.
l.)
Close the top cover again.
3.10
Change of Motor
Required tools:
1.5 mm Allen Key
2.5 mm Allen Key
4 mm Allen Key
8 mm open-ended wrench
a.)
On the RHS unscrew the 2 x 4mm allen screws securing the purge valve module to
gain better access to the motor.
Fig. 3.10.1: RHS – Unscrewing purge valve module