2009 by ELTRA GmbH Germany – July 2009 – Service Manual Thermostep
Page 35
d.)
In case of a gas leakage remove the corresponding feeding gas tube out of its
connector and insert it again.
e.)
If it is not possible to correct the gas leckage by this remove the feeding gas tube out
of the connector again. Check with the open-ended wrench if the connector is fixed
completely.
f.)
Replace the connector if necessary.
g.)
Is a manual switch of the valve via the button not possible it has to be replaced. Note
the orientation of the feeding gas tubes and remove them from their connectors. (By
pressing the coloured ring at the vent of the connector you can remove the tube!)
Fig. 3.2.3: LHS – Marking of the tubes
h.)
The power supply of the valve is fixed by a screw (transparent header). Unfasten this
screw (underside of the valve).
Fig. 3.2.4: LHS – Unfasten the power supply of the valve