MAINTENANCE
22
NP-DK50 DI-15_08-2020
Check tightness of joints and inspect the equipment
Leakage test:
Turn on the compressor
Stop consumption of compressed air.
If the pneumatic system does not leak, the compressor must at a short
time (within 5 minutes) turn off.
If the compressor is not turned off, the pneumatic system is leaking. Use
soapy water to find the leaky joint. Start with the joints outside the case,
and then test the ones inside the case after removing the covers.
Tighten or re-seal joints as necessary.
Inspect the equipment:
Check the status of the compressor motor for balanced operation or noise.
Check the condition of the hangers above the pump.
Check fan functionality.
Check the condition of the filter. With regular operation, condensate from
the filter drains automatically.
Check the status of air pump
If necessary, replace any faulty parts.
Replacing the suction filter
Unlock the cover and pull it out.
Replace the filter, see Chapter
Maintenance materials
Lock the cover
Replacing of air filter cartridge
Unscrew the two screws on the service cover below the filter vessel,
remove the cover.
Remove the vessel and replace the filter cartridge, see Chapter
Maintenance material.
Screw the filter vessel, screw on the service cover.
Replacing of fuses
The fuse failure is usually result by failure on the device. Before replacing
a fuse, fix failure on device.
Summary of Contents for invent DK50 DI
Page 1: ...User manual EN Manuel de l utilisateur FR N vod na pou itie SK RU N vod k pou it CS...
Page 2: ......
Page 62: ...C 62 NP DK50 DI 15_08 2020...
Page 63: ...C NP DK50 DI 15_08 2020 63 e Caution risk of electric shock...
Page 64: ...C 64 NP DK50 DI 15_08 2020 Manufacturer IP21 Te...
Page 65: ...C NP DK50 DI 15_08 2020 65 OUT 7 4...
Page 66: ...66 NP DK50 DI 15_08 2020 2 1 2 3 4 USB 5 6 7 8 Ethernet 9 10 11 12 13 14 15 16 17 18...
Page 67: ...NP DK50 DI 15_08 2020 67 BAR PSI E O AC OC 1 7 5 0 65 C 15...
Page 68: ...68 NP DK50 DI 15_08 2020 AUTO OFF 2 8 AUTO ON 2 5 POWER SAVE FULL OFF 15...
Page 69: ...NP DK50 DI 15_08 2020 69 DK50 DI STANDBY DK50 DI...
Page 72: ...72 NP DK50 DI 15_08 2020 4...
Page 73: ...NP DK50 DI 15_08 2020 73 5...
Page 74: ...74 NP DK50 DI 15_08 2020 WALL WALL OUT OUT OUT WALL 5 5 80...
Page 75: ...NP DK50 DI 15_08 2020 75 I I STANDBY RUN...
Page 77: ...NP DK50 DI 15_08 2020 77 5 o o e...
Page 78: ...78 NP DK50 DI 15_08 2020 5 a...
Page 80: ...80 NP DK50 DI 15_08 2020 9 031500034 000 5...
Page 81: ...NP DK50 DI 15_08 2020 81...
Page 83: ...NP DK50 DI 15_08 2020 83 6 C WALL C SERVICE...
Page 84: ...84 NP DK50 DI 15_08 2020 7 1 N PE 100 230V 50 60 Hz I PCB XC1 FU1 CB1 M1 YV1 EV1 ST...
Page 85: ...NP DK50 DI 15_08 2020 85 CO PP DR DV EV AC RV SV P1 P2...