ELECTROMAGNETIC COMPATIBILITY DECLARATION
NP-DK50 DI-15_08-2020
29
Guidance and manufacturer's declaration - electromagnetic immunity
Pursuant to IEC 60601-1-2:2014 - Medical electrical equipment - Part 1-2: General requirements for
basic safety and essential performance - Collateral Standard: Electromagnetic disturbances
The equipment is intended for use in the electromagnetic environment specified below. The customer or the
user of the equipment should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply
lines
±1 kV for
input/output lines
±2 kV
100 kHz repetition
frequency
Applied to mains
connection
Mains power quality should be
that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV L-N
±2 kV L-PE; N-PE
Applied to mains
connection
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short interruption,
and voltage
variations on
power
supply
input lines IEC
60601-4-11
U
T
=0%, 0,5 cycle
(at 0, 45, 90, 135,
180, 225, 270 and
315°)
U
T
=0%, 1 cycle
U
T
=70% 25/30
cycles (at 0°)
U
T
=0%, 250/300
cycles
U
T
=>95%, 0,5 cycle
(at 0, 45, 90, 135, 180,
225, 270 and 315°)
U
T
=>95%, 1 cycle
U
T
=70% (30% dipp U
T
),
25(50Hz)/30(60Hz)
cycles (at 0°)
U
T
=>95%,
250(50Hz)/300(60Hz)
cycle
Mains power quality should be
that of a typical commercial or hospital
environment.
The device stops and restarts
automatically at each dip. In this case
does not occur unacceptable pressure
drop.
Power frequency
(50/60 Hz)
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment
NOTE: U
T
is the A.C. mains voltage prior to application of the test level.
Summary of Contents for invent DK50 DI
Page 1: ...User manual EN Manuel de l utilisateur FR N vod na pou itie SK RU N vod k pou it CS...
Page 2: ......
Page 62: ...C 62 NP DK50 DI 15_08 2020...
Page 63: ...C NP DK50 DI 15_08 2020 63 e Caution risk of electric shock...
Page 64: ...C 64 NP DK50 DI 15_08 2020 Manufacturer IP21 Te...
Page 65: ...C NP DK50 DI 15_08 2020 65 OUT 7 4...
Page 66: ...66 NP DK50 DI 15_08 2020 2 1 2 3 4 USB 5 6 7 8 Ethernet 9 10 11 12 13 14 15 16 17 18...
Page 67: ...NP DK50 DI 15_08 2020 67 BAR PSI E O AC OC 1 7 5 0 65 C 15...
Page 68: ...68 NP DK50 DI 15_08 2020 AUTO OFF 2 8 AUTO ON 2 5 POWER SAVE FULL OFF 15...
Page 69: ...NP DK50 DI 15_08 2020 69 DK50 DI STANDBY DK50 DI...
Page 72: ...72 NP DK50 DI 15_08 2020 4...
Page 73: ...NP DK50 DI 15_08 2020 73 5...
Page 74: ...74 NP DK50 DI 15_08 2020 WALL WALL OUT OUT OUT WALL 5 5 80...
Page 75: ...NP DK50 DI 15_08 2020 75 I I STANDBY RUN...
Page 77: ...NP DK50 DI 15_08 2020 77 5 o o e...
Page 78: ...78 NP DK50 DI 15_08 2020 5 a...
Page 80: ...80 NP DK50 DI 15_08 2020 9 031500034 000 5...
Page 81: ...NP DK50 DI 15_08 2020 81...
Page 83: ...NP DK50 DI 15_08 2020 83 6 C WALL C SERVICE...
Page 84: ...84 NP DK50 DI 15_08 2020 7 1 N PE 100 230V 50 60 Hz I PCB XC1 FU1 CB1 M1 YV1 EV1 ST...
Page 85: ...NP DK50 DI 15_08 2020 85 CO PP DR DV EV AC RV SV P1 P2...