OPERATION
NP-DK50 DI-15_08-2020
17
Compressed air connection
Air inlet (WALL)
Connect the compressed air line from the central distribution line to the
WALL quick coupling - inlet compressed air (if a central air distribution line is
available).
Air outlet (OUT)
Connect the pressure hose to the medical ventilator to the OUT quick
coupling on the compressed air outlet.
Air from the central distribution system is automatically connected through
the compressor to the OUT outlet port. In this configuration, the compressor
serves as a backup source of compressed air. If the air pressure from the
central distribution system is reduced, the compressor automatically switches
on and there is no interruption in the supply of continuously pressurized air at
the outlet of the compressor.
The maximum rated air pressure for the WALL quick coupling is 5.5 bar
/ 80 psi. Please note that the air supply from the central distribution line
entering the compressor must be medical grade air (particulate size
and moisture content). The compressor does not modify the air from
the central distribution line.
The hose connecting the compressor to medical ventilator must not
pass through a cold environment i.e. placed on the ground. It should be
as short as possible with no kinks (this may cause water to condense
inside the hose).
The compressor must not be used for supplying air to the medical air
central distribution pipeline system.
Electrical connection
The compressor comes with a plug containing an appropriate
protective contact (ground) adhere to local electrical regulations. The
voltage and frequency of the mains must comply with the
specifications on the data label.
The electrical cord must not be stressed or have any tension exerted
upon it, and must always be free.
The socket must be accessible for safety reasons so that the equipment
can be safely disconnected from the power supply in case of an
emergency.
The relevant current circuit must be protected.
Summary of Contents for invent DK50 DI
Page 1: ...User manual EN Manuel de l utilisateur FR N vod na pou itie SK RU N vod k pou it CS...
Page 2: ......
Page 62: ...C 62 NP DK50 DI 15_08 2020...
Page 63: ...C NP DK50 DI 15_08 2020 63 e Caution risk of electric shock...
Page 64: ...C 64 NP DK50 DI 15_08 2020 Manufacturer IP21 Te...
Page 65: ...C NP DK50 DI 15_08 2020 65 OUT 7 4...
Page 66: ...66 NP DK50 DI 15_08 2020 2 1 2 3 4 USB 5 6 7 8 Ethernet 9 10 11 12 13 14 15 16 17 18...
Page 67: ...NP DK50 DI 15_08 2020 67 BAR PSI E O AC OC 1 7 5 0 65 C 15...
Page 68: ...68 NP DK50 DI 15_08 2020 AUTO OFF 2 8 AUTO ON 2 5 POWER SAVE FULL OFF 15...
Page 69: ...NP DK50 DI 15_08 2020 69 DK50 DI STANDBY DK50 DI...
Page 72: ...72 NP DK50 DI 15_08 2020 4...
Page 73: ...NP DK50 DI 15_08 2020 73 5...
Page 74: ...74 NP DK50 DI 15_08 2020 WALL WALL OUT OUT OUT WALL 5 5 80...
Page 75: ...NP DK50 DI 15_08 2020 75 I I STANDBY RUN...
Page 77: ...NP DK50 DI 15_08 2020 77 5 o o e...
Page 78: ...78 NP DK50 DI 15_08 2020 5 a...
Page 80: ...80 NP DK50 DI 15_08 2020 9 031500034 000 5...
Page 81: ...NP DK50 DI 15_08 2020 81...
Page 83: ...NP DK50 DI 15_08 2020 83 6 C WALL C SERVICE...
Page 84: ...84 NP DK50 DI 15_08 2020 7 1 N PE 100 230V 50 60 Hz I PCB XC1 FU1 CB1 M1 YV1 EV1 ST...
Page 85: ...NP DK50 DI 15_08 2020 85 CO PP DR DV EV AC RV SV P1 P2...