Display
Meaning
Effect
Cause
Remedy
No communication
between dongle box
and pump
pump stops
Transmission of the data
has been disrupted
during operation
Check the connection
and replace the dongle
box if necessary
Operating faults: Alarm messages (display) Fault 12
No communication
between MicroFlow
box and pump
pump stops
Transmission of the data
has been disrupted
during operation
Check the connection
and replace the
MicroFlow if necessary
Operating faults: Alarm messages (display) Fault 13
12.2.1 Disabling dongle box, OGM
PLUS
, MicroFlow
PLUS
in the pump software
The following procedure is applied to re-disable a peripheral unit
(dongle- box, OGM
PLUS
, MicroFlow
PLUS
), which had previously been connected and was
then removed again, in the pump software:
Cancel the alarm signal which appears (see fig. 12.10, 12.11, 12.12) by pressing the Start/
Stop button.
Es erscheint folgende Anzeige im Display:
Dongle box
OGM
PLUS
MicroFlow
PLUS
Disabling dongle box in the
pump software
Disabling OGM
PLUS
in the pump
software
Disabling MicroFlow
PLUS
in the pump
software
Deactivation can be selected by pressing the
s
or
t
button, and the following is displayed
Dongle box
OGM
PLUS
MicroFlow
PLUS
Disabling dongle box in the
pump software
Disabling OGM
PLUS
in the pump
software
Disabling MicroFlow
PLUS
in the pump
software
The setting is applied by pressing the Start/Stop button.
Operating faults
114
417101442 Rev. 12-02.2019
Summary of Contents for Elados EMP KKS E60
Page 2: ......
Page 123: ...14 2 Abmessungen Abb 89 Abmessungen Technische Daten 121 417101442 Rev 12 02 2019 ...
Page 135: ......
Page 258: ...14 2 Dimensions Fig 90 Dimensions Technical Specifications 122 417101442 Rev 12 02 2019 ...
Page 393: ...14 2 Dimensions Fig 93 Dimensions Données techniques 123 417101442 Rev 12 02 2019 ...
Page 405: ......