10.3.14.6
MicroFlow
PLUS
/ Metering delay
Use the "Metering delay" option to specify the time during which the metering monitoring
function is disabled when handling out gassing media to allow time for degassing the
system.
Once the metering pause time has expired, the "Metering delay" is activated as soon as
the next metering is initiated
(
Fig. 69: Configuration MicroFlow
PLUS
/ Metering delay
Example:
Metering pause = 240 s,
Metering delay = 40 s,
This example has a metering pause of 240 seconds during which metering is stopped.
The metering delay of 40 seconds is initiated with the next metering request during which
the metering monitoring of the MicroFlow
PLUS
is interrupted.
Any gas bubbles can be removed from the system without triggering an alarm during this
time.
Setting range: 2 - 3600 s
Default setting = 40 sec.
Menu description
92
417101442 Rev. 12-02.2019