![Dräger 84 13 050 Operating Instructions Manual Download Page 59](http://html.mh-extra.com/html/drager/84-13-050/84-13-050_operating-instructions-manual_2526714059.webp)
Gas Supply
The supply gas (O
2
, Air, or mixture as required) pro-
vided by an oxygen blender is cleaned through filter
1
and reduced to a constant pressure with pressure
regulator
2
. The 3/2-way solenoid valve »Insp./Exp.«
3
releases inspiratory flow in the rhythm of the set venti-
lator rate. The electrically operated flow valve
10
controls the flow.
Supply gas is also routed to 3/2-way solenoid valve
»Insp./Exp.« and to demand valve
6
via 3/2-way sole-
noid valve »CMV/CPAP«. System pressure is supplied
to PEEP valve
8
via solenoid valve
7
and is monitored
by pressure sensor
9
.
CMV/SIMV
Inspiration
3/2-way solenoid valve »Insp./Exp.«
3
releases gas flow.
Gas flow is controlled via electrically operated flow
valve
10
and reaches the patient circuit via check valve
13
, humidifier
11
, wye connector
12
, and flow sensor
15
. At the same time, expiratory valve
14
is closed. The
controlled 2/2 way valve
21
prevents the inspiratory gas
from escaping to the mechanical PEEP valve
8
. The
pressure-limiting valve
22
limits the
inspiratory pressure to a maximum value, independent
of the P
max
control unit.
Theory of Operation
Symbols for Pneumatic Components
Theory of Operation
Symbols for Pneumatic Components
Gas Supply
CMV/SIMV
59
Operating InstructionsMicroVent
P
E
∆
P
E
Filter
Check valve
Pressure regulator
Pressure limiting valve, adjustable
Pressure limiting valve with fixed setting
2/2-way valve, pneumatically controlled
3/2-way valve, electrically controlled
Pressure gauge
Pressure sensor
Differential pressure sensor
Expiratory valve, pneumatically-controlled
P
E
E
P
P
E
∆
P
E
PEEP
Paw
Paw
Flow
24
23
22
6
25
26
5
IPPV/
SIMV/CPAP
4
21
13
11
12
15
Patient
14
8
7
10
O
2
AIR
Gasmischer
O
2
/ AIR
1
2
3
9
Insp./Exsp.
27
28
ASB
29
O2/Air
Blender
Press.
Control
CMV
SIMV/CPAP
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...