Care
-
Always clean and disinfect expiration valve and
(reusable) patient circuit after each use.
-
Clean ventilator and high pressure hoses according
to visual check of cleanliness and/or established
hospital procedures.
Disassembly
1
Remove pressure sensing hoses (flow measurement)
from flow sensor.
2
Pull elbow connector from flow sensor.
3
Disconnect flow sensor from wye.
●
Empty water traps in patient circuit.
●
Disconnect patient circuit from nozzles
CAUTION !
When removing patient circuits, always handle at
end fittings! Otherwise, support helix may be torn at
end fitting or hose and fitting could come apart.
CAUTION !
Prevent any foreign objects from getting into the
flow sensor.
Do not use pressurized air for cleaning, the orifice
membrane inside the flow sensor may become dam-
aged.
WARNING !
Always follow accepted hospital procedures for
handling equipment contaminated with body
fluids.
Disassembly
Care
42
Operating Instructions MicroVent
2
3
1
1
Summary of Contents for 84 13 050