![Dräger 84 13 050 Operating Instructions Manual Download Page 16](http://html.mh-extra.com/html/drager/84-13-050/84-13-050_operating-instructions-manual_2526714016.webp)
1
Attach flow sensor to wye.
2
Push elbow connector onto flow sensor.
When using a bacterial filter
●
Always connect bacterial filter to elbow connector as
shown.
●
Push pressure sensing lines for flow measurement
onto flow sensor nipples – note different diameters of
hose ends and respective nipples.
Pneumatic Connections
1
Attach gas supply line and tighten.
2
Attach pressure sensing lines for flow measurement
to connector on MicroVent.
CAUTION !
Do not pressurize this connector for pressure
sensing lines. It might destroy the internal sensor.
CAUTION !
Always use elbow connector for accurate flow
measurement
Preparation
Assembly
16
Operating Instructions MicroVent
1
2
4
3