Continuous Positive Airway Pressure
(CPAP) / Pressure Support
Checking for sufficient spontaneous breathing on
pressure gauge:
Patient must be capable of generating a negative
pressure of at least 2 cmH
2
O during inspiration.
1
Set knob »f« to 0.
2
Set switch for ventilator modes to SIMV/CPAP.
3
Set knob »PEEP« to desired level of positive airway
pressure.
NOTE:
When set values are below 3 cmH
2
O, a
disconnection alarm is generated. Disconnections are
detected when pressure falls below the absolute thres-
hold of 3.0 cmH
2
O or when there is a sudden drop in
pressure of more than 5 cmH
2
O during CPAP.
For pressure support:
4
set knob »PS« higher than PEEP, as
appropriate. Pressure support will then be active
according to the difference in pressure levels between
Press. Supp. and PEEP.
NOTE:
When Press. Supp. is set lower than PEEP, no
pressure support is provided
If apnea occurs, apnea ventilation is switched on at
f = 8 bpm:
5
Use knob »V
T
« to set tidal volume and
6
Use knob »T
insp
« to set inspiratory time.
When no breaths have been detected by the ventilator for
15 seconds, display shows warning message:
continuous audible alarm begins,
red alarm light starts flashing.
If ventilator does not detect an inspiratory effort after
another 15 seconds, apnea ventilation will begin - using a
fixed ventilator rate of 8 bpm and the values set for V
T
and T
insp
. The display will then show:
Apnea ventilation
Apnea
WARNING !
Use CPAP mode only for patients who have suffi-
cient spontaneous breathing.
Continuous Positive Airway
Pressure (CPAP) / Pressure
Support
Operation
34
Operating Instructions MicroVent
D
3
4
1
6
5
2
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...