The table below is intended as a quick reference for
recognizing and correcting causes of alarms.
NOTE:
The messages in the table are listed in
alphabetical
order.
Troubleshooting
Alarm screen messages are displayed hierarchically,
i.e., in order of priority.
If, for example, two faults are detected simultaneously,
the more serious fault will be displayed.
Troubleshooting
50
Operating Instructions MicroVent
Problem / Message
Likely causes
Remedy
AC power failure
No external supply available.
Connect to electrical power supply
as soon as possible.
Apnea
Alarm status light flashing,
intermittent audible alarm.
Patient's spontaneous breathing in
CPAP mode insufficient.
Switch to CMV or SIMV.
Bat. discharged
Alarm status light flashing,
intermittent audible alarm.
Battery discharged; no external
power supply.
Replace battery pack with new pack
or connect to external power supply.
Change bat.
Alarm status light flashing, single
beep every 30 seconds. Display
illumination cannot be switched on.
Battery pack will be discharged
within a few minutes, external power
supply not connected.
Replace battery pack with new pack
or connect to external power supply.
Charge NiCd
Alarm status light flashing, single
beep every 30 seconds. Display
illumination cannot be switched on.
NiCd pack will be discharged within
a few minutes, external power supply
not connected.
Replace rechargeable battery pack
with a fully charged NiCd pack or
connect to external power supply.
Check settings
Alarm status light flashing,
intermittent audible alarm.
Settings utside ventilator performance
limits; required flow is less than
10 L/min or greater than 80 L/min.
Adjust set parameters; e.g.
f, V
T
or T
insp.
Defective NiCd
Ventilator connected to external
power supply. Defective NiCd pack.
Replace rechargeable battery pack
with a new NiCd pack.
Flow measurement INOP
Flow measurement not operable.
Signals from flow measurement are
not used for alarms.
Ventilation may be continued.
After shutdown, call DrägerService.
Leakage
Alarm status light flashing,
intermittent audible alarm.
The measured expiratory minute
volume has fallen to about 40 % be-
low inspiratory minute volume.
Correct leakage in patient system
and/or in ET-tube, if necessary.
Loss of CPAP
CPAP drops suddenly by more than
5 cmH
2
O or
circuit connections leaking.
Change CPAP pressure in small
increments only. Make sure
connections are leak free.
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...