Inspection and
Maintenance
Preventive Maintenance
45
Operating Instructions MicroVent
Inspection and Maintenance
Preventive Maintenance
Refer to schedule to the right for preventive maintenance
intervals:
For exchanging the NiCd battery pack, please refer to
pages 17-18, "Preparation".
WARNING !
Treatment of batteries:
Do not throw into fire! Do not force open!
danger of bodily injury.
Follow all local, state, and federal regulations
with respect to environmental protection when
disposing of batteries.
WARNING !
Preventive Maintenance work on the MicroVent
ventilator may be performed by trained and
factory authorized staff only.
WARNING !
To avoid any risk of infection, clean and disinfect
ventilator and accessories before any mainte-
nance according to established hospital proce-
dures - this applies also when returning ventila-
tors or parts for repair.
Scheduled safety checks
every 6 months
Inspection and prev. maint.
every 6 months
Internal alkaline batteries
replace, if message
»Bat. discharged« or
»Change bat.«
is displayed
Internal NiCd battery pack
replace, if message
»defective NiCd« is dis-
played, every 2 years at
the latest
Pressure reducer overhaul
every 6 years
Gas mixer overhaul
Service after 3 years by
trained service per sonnel
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...