When external voltage is connected:
•
the green status light = "DC power available"
will be on, independent of whether the ventilator is
turned »On« or »Off«. The internal battery pack is
now being charged.
•
Display while ventilator is turned off:
•
It takes about 5 hours to recharge a completely
discharged NiCd battery pack.
•
Display while ventilator is turned off:
At room temperature, the ventilator can be operated
for about 2 hours with a fully charged NiCad battery
pack.
•
Near the end of battery capacity,
the ventilator displays
Approximately 10 minutes of operation remain at
the time of this message.
NOTE: Low ambient temperatures and the actual
condition of the rechargeable batteries may reduce
operating times.
Use with external DC supply
Voltage fluctuations in on-board supplies can be sig-
nificant, e.g. due to varying load conditions. They may
exceed the input tolerances specified for MicroVent .
The DC/DC converter conditions these types of sup-
plies to provide a constant voltage of approximately
12 V DC.
•
Attach cigarette lighter adapter plug to DC source,
plug other end into external DC input jack on top
of ventilator.
CAUTION !
Always use DC/DC converter 84 12 017 when
using a DC source.
Charge NiCd
Stand-by
NiCd charged
Stand-by
NiCd charging
Electrical Supply
Preparation
20
Operating Instructions MicroVent
D
1
12 V DC
24 V DC
28 V DC
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...