14 Liquid crystal display for minute volume and other
measured parameter values, for alarms and advisory
messages.
15 Knob for ventilator rate »f«, continuously adjustable
from 5 to 60 bpm in CMV and from 2 to 60 bpm in
SIMV. Setting of 0 bpm for CPAP.
16 Key for switching on display backlighting for
30 seconds and for checking displays and audible
alarm. Also for switching between "Info" functions
displaying measured values of parameters flow,
I:E-ratio, pressure, rate, V
Texp
.
What's What
Front view
1
Pressure gauge for airway pressure
2
Knob for pressure limitation »P
max
«; continuously
adjustable from 20 to 80 cmH
2
O
3
Double function knob for upper and lower alarm
limits for minute volume
4
Knob for Pressure Support »PS«, continuously
adjustable from 0 to 35 cmH
2
O
5
Knob for »PEEP/CPAP«, continuously adjustable
from 0 to 18 cmH
2
O
6
On/off switch »ON/OFF«
7
Switch for CMV/AC and SIMV/CPAP ventilation
modes
8
Knob for setting inspiratory time, T
insp
., between
0.2 and 3 seconds
9
Yellow status light »Using internal supply«
10 Green status light »External DC power supply con-
nected«
11 Red alarm status light
12 Key for silencing audible alarms for 2 minutes and for
resetting screen alarm messages
13 Knob for tidal volume, V
T
, continuously adjustable
from 0.1 L to 1.5 L
Front View
What's What
53
Operating Instructions MicroVent
2-679-95
1
15
16
14
13
12
11
10
9
7
2
3
4
5
6
8
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...