Operating Data
Gas supply
Supply gas
O
2
, medical air, or mixtures
O
2
concentration
±
3 % of set value when
operated with blender
84 12 411
Requirements for the
medical grade from wall
supply gas
pipeline, from cylinders, or
from air compressors
Supply pressure
38 to 84 psi (2.7 to 6.0 bar)
at 80 L/min.
Max press. difference
between O
2
and Air when
operated with blender:
20 psi (1.2 bar)
Gas connections
pursuant ISO 5359, Annex A
(DISS)
Gas consumption
Ventilator
for internal pneumatic
control circuits
approximately 1.0 L/min
about 2.5 L/min base flow in
any ventilation mode (TI:TE 1:1)
Blender
approximately 12 L/min
O
2
-Air mixture
Nebulizer port
approximately 3 L/min
at 72 psi (5 bar) supply
pressure
Typical times of operation
with minute ventilation
of 10 L/min
without blender:
E-cylinder
approximately 46 minutes
WARNING !
Always use medical grade oxygen or air that is
dry and free from dust and oil. Contaminated gas
may cause ventilator malfunction.
WARNING !
MicroVent is not intended to be used with gas
mixtures of nitrous oxide
Alarms
Supply press. low
Alarm is activated when
supply pressure falls below
approx. 22 psi (1.5 bar).
Paw high
Adjusted with knob »Pmax«,
alarm is activated when set
value of Pmax is reached.
Paw low
Alarm is activated when no air-
way pressure difference of at
least 5 cmH
2
O has been
reached within two ventilator
breaths in CMV or SIMV or if
pressure falls below 3 cmH
2
O in
CPAP mode.
PEEP high
Alarm is activated when airway
pressure don´t falls below
22 cmH
2
O during expiration.
Leakage
Alarm is activated when the ex-
piratory minute ventilation is less
than 60 % of inspiratory minute
ventilation. The »Leakage« alarm
is not functional in CPAP mode.
Apnea
Only functional in CPAP mode.
Alarm is activated when no
breaths have been detected for
15 seconds.
Check settings
Alarm is activated when an in-
spiratory flow outside the range
of 10 to 80 L/min would result
from the set combination of V
T
,
Rate, and Tinsp. It is not a mes-
sage about an operator error
and does not need to be reset.
Alarms
are generated as visual and audi-
ble alarms.The audible portion
of an alarm is muted after a con-
dition has been corrected. The
message displayed must be re-
set (to acknowledge).
Self test
automatically in fixed time intervals
during operation and standby.
Audible alarm loudness 75 dB (A) in 1 m distance
high rate
Alarm is activated when the
breathing rate exeeds 30/min
during 3 minutes.
This alarm can switched
ON/OFF in the Customer
Service Mode.
Alarms
Operating Data
Technical Data
57
Operating Instructions MicroVent
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...