Preparation
Installing the Patient Valve
1
Insert silicone membrane into open patient valve -
watch for correct orientation of membrane
2
Put on lid and turn clockwise 90o to lock.
3
Push expiratory valve into the opening on the
ventilator = nozzle pointing downwards.
4
Screw or push O
2
sensor firmly into the O
2
sensor
adapter.
5
Attach O
2
sensor adapter to nozzle.
●
Connect ventilator circuit to nozzles.
Note suggested length of circuit segments
(in meters).
●
Make sure watertraps are positioned vertically.
●
Prepare humidifier according to its Operating
Instructions.
WARNING !
Do not use antistatic (electrically conductive)
patient circuits or tubing.
WARNING !
Always use an independent oxygen analyzer to
monitor inspiratory oxygen concentrations.
WARNING !
Always verify that valve membrane is installed in
correct orientation. Ventilator malfunction will
occur if silicone membrane is reversed.
Take care not to damage or bend rubber disc
inside patient valve. Ventilator malfunction will
occur.
Assembly
Preparation
15
Operating Instructions MicroVent
2
1
3
D
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...