O
2
Supply from an oxygen cylinder
Always use full tank (2400 psi)
• Attach pressure regulator for delivery pressure of 38
to 84 psi ( = 2.7 to 6.0 bar), nominal pressure 50 psi
to oxygen cylinder.
• Connect MicroVent with DISS high pressure hose to
oxygen regulator.
• Open valve
slowly
and completely.
Determining gas supply in O
2
cylinder
Example:
Tank pressure according to pressure gauge
on regulator:
Supply of pressurized O
2
(E-cylinder):
Estimate for expected time of operation of
MicroVent
Example:
Mode CMV:
Minute Ventilation:
expiratory gas consumption due to the base flow
pressur. gas supply [L]
Expected time of op. =
(MV + 1) + 1 – Tinsp. · f · 5[L/min]
60
•
Close cylinder valve completely after test
Supply with oxygen from wall pipeline system
•
Attach high pressure hose with DISS connector to
MicroVent and connect to wall terminal with
appropriate quick connect probe.
_______
* Internal ventilator gas consumption approximately 1 L/min
Gas Supply
Preparation
22
Operating Instructions MicroVent
Rate = 10/min, V
T
= 1 L, Tinsp. = 3 s
MV = 10/min x 1 L = 10 L/min
= 1 – Tinsp. · f · 5 [L/min]
60
T
op
= 622 / 13,5 = 46 minutes
2400 psi (165 bar)
622 L
O
2
(
)
(
)
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...