![Dräger 84 13 050 Operating Instructions Manual Download Page 26](http://html.mh-extra.com/html/drager/84-13-050/84-13-050_operating-instructions-manual_2526714026.webp)
Testing O
2
Concentration
using an independent, external oxygen analyzer
1
Set mixer to 50 vol.% O
2
.
●
Measured O
2
concentration should be in the
range of 50
±
5 vol.%.
Testing Minute Volume Monitoring
1
Set lower alarm limit to 7 L/min.
●
Display and intermittent audible alarm
1
Reset lower alarm limit to 4 L/min. The intermittent
audible alarm should stop
To clear display:
2
Press
g
key.
1
Set upper alarm limit to 5 L/min.
●
Display and intermittent audible alarm
1
Reset upper alarm limit to 8 L/min. The intermittent
audible alarm should stop
To clear display:
2
Press
g
key.
CMV MV = 6.0
MV high
CMV MV = 6.0
MV low
Testing O
2
Concentration
Testing Minute Volume
Monitoring
Tests of Readiness for
Operation
26
Operating Instructions MicroVent
D
1
D
1
2