In Case of Loss of Electric Power
NOTE: In case of an AC power failure (blackout),
automatic ventilation, volume measurement and
alarms continue to operate.
If this situation has occurred, ventilator display shows:
●
Acknowledge = press
g
key.
The ventilator can be operated for up to 2 hours without
an electrical power supply.
●
Make sure than an electrical power supply is
re-established in good time.
WARNING !
Automatic patient ventilation, volume measure-
ment, and alarms are not functional in case of a
total loss of electric power!
Spontaneous breathing through the integrated
demand valve, however, is still possible in that
situation,
Immediately start ventilating patient with inde-
pendent manual ventilation device (resuscita-
tion bag) using PEEP and/or increased inspirato-
ry oxygen concentration where necessary and
appropriate.
main supply down
Operation
In Case of Loss of
Electric Power
38
Operating Instructions MicroVent