Continuous audible alarm and flashing of red alarm
status light both continue.
When spontaneous breathing is insufficient:
●
switch to SIMV or CMV.
To switch off apnea ventilation:
●
press
g
key twice.
Pressure-limited ventilation
Use optional pressure limiting valve 8405390!
●
Draw ventilation hose off twin nozzle of the
humidifier.
●
Draw elbow connector (with twin nozzle) off the
humidifier connection.
1
Plug twin nozzle (with elbow connector) into the
humidifier connection.
2
Plug pressure limiting valve onto elbow connector
with adjusting screw pointing upwards.
3
Plug ventilation hose onto nozzle of pressure limiting
valve.
●
Set ventilation as described for IPPV/SIMV mode.
●
Set inspiratory pressure by means of adjusting screw
on pressure limiting valve and check on pressure
gauge.
●
Check set tidal volume V
T
.
●
Correct alarm limits for the minute volume MV and
max. airway pressure Paw if necessary.
Operation
Pressure-limited ventilation
35
Operating Instructions MicroVent
2
1
3