➄
On/off switch, ON/OFF
➅
Pressure gauge
for continuous display of airway
pressure
➆
Liquid crystal display
for
–
measured parameter values
(MV, PEEP,
Peak, Mean, I:E, Insp. flow)
–
Alarms
and
advisory messages
with
l
key to switch on display backlighting
and to display measured parameter values.
➇
g
key
to temporarily silence audible alarms and
to reset alarm messages after correcting the cause
of an alarm.
Overview
Front Panel
➀
The knobs in the center of the front panel set:
–
ventilator rate
f
–
tidal volume
V
T
–
inspiratory time
T
insp
.
➁
Below, grouped together in one field, are the
knobs for:
–
max. airway pressure P
max
Y
–
minute volume alarm limits, MV
Y
➂
Next to this field, the knob for settting
–
pressure support
during spontaneous
breathing
–
endexpiratory pressure, PEEP
and positive
airway pressure,
CPAP
➃
Next to it, the switch for
ventilation modes
–
CMV
/ A/C
or
SIMV / CPAP
Overview
Front Panel
14
Operating Instructions MicroVent
2-679-95