![Dräger 84 13 050 Operating Instructions Manual Download Page 44](http://html.mh-extra.com/html/drager/84-13-050/84-13-050_operating-instructions-manual_2526714044.webp)
Bath disinfection:
Use for disassembled parts of the patient valve, flow sen-
sor, ventilator circuit (where applicable), pressure sen-
sing hoses for flow measurement.
●
Disinfect by immersion in disinfectant solution, follo-
wing manufacturer's instructions. Move parts tho-
roughly.
●
Remove parts after the recommended exposure time
and rinse thoroughly with distilled water.
●
Shake water out and let parts dry completely before
reassembling.
Sterilization procedures
Where referenced in established hospital procedures
●
Disassembled parts of the patient valve, flow sensor,
ventilator circuit (where applicable), pressure sensing
hoses for flow measurement may be steam autocla-
ved at 273 ºF (134 ºC).
Accessoroes
Cleaning and disinfection procedures for humidifier and
oxygen analyzer should be performed according to
their respective Operating Instructions
After cleaning and disinfecting
●
Re-assemble (page 15),
●
connect power and gas supply (pages 17,21),
●
check readiness for operation (page 23).
WARNING !
Water remaining in the ventilator valve or in the
pressure sensing lines for flow measurement
may cause ventilator malfunction.
CAUTION !
Do not use hard brush for cleaning
Prevent any foreign objects from getting into the
patient valve.
Disinfecting / Sterilizing
To prevent any damage, we recommend that only deter-
gents and disinfectants are used that are compatible with
the materials used in the ventilator and its accessories,
e.g. surface disinfectants on the basis of aldehydes,
alcohol, or quarternary ammonium compounds for
disinfection of the main unit.
Ensure that all disinfectants are registered with the U.S.
Environmental Protection Agency for use as intended.
Always follow the instruction labels specifically with
respect to prescribed concentrations and the necessary
exposure times.
Wipe Disinfection:
Use for ventilator, circuit support arm, mobile stand, and
high pressure hoses.
●
Disinfect by wiping all surfaces following manufac-
turer's instructions of disinfectant used regarding
concentrations and application.
CAUTION !
Follow all accepted hospital procedures for disin-
fecting parts contaminated by body fluids (protec-
tive clothing, eyewear, etc.).
CAUTION !
Certain components of the ventilator consist of
materials that are sensitive to certain organic sol-
vents sometimes used for cleaning and disinfec-ting
(e.g., phenols, halogen releasing compounds, oxy-
gen releasing compounds, strong organic acids,
etc.). Exposure to such substances may cause
damage that is not always immediately recognized.
Sterilization of the ventilator itself with ethylene ox-
ide (EtO) is also not recommended.
Disinfecting/Sterilizing
Care
44
Operating Instructions MicroVent
Summary of Contents for 84 13 050
Page 2: ...Operating Instructions MicroVent This page intentionally left blank ...
Page 4: ...4 Operating Instructions MicroVent This page intentionally left blank ...
Page 13: ...Overview 13 Operating Instructions MicroVent 2 682 95 ...
Page 65: ...This page intentionally left blank 65 ...
Page 66: ...This page intentionally left blank 66 ...