Inspect the system
WARNING
Do not exceed the top shelf weight limit of 34 kg (75 lb).
w
Make sure that the breathing circuit is correctly connected
and not damaged. Replace the breathing circuit if it is
damaged.
w
Do not leave gas cylinder valves open if the pipeline
supply is in use. Cylinder supplies could be depleted,
leaving an insufficient reserve supply in case of pipeline
failure.
Before using the system, make sure that:
•
The equipment is not damaged.
•
Components are correctly attached.
•
The breathing circuit is correctly connected, not damaged, and
the breathing system contains sufficient absorbent.
•
The vaporizers are locked in position and contain sufficient agent.
•
Pipeline gas supplies are connected and the pressures are
correct.
•
Cylinder valves are closed on models with cylinder supplies.
•
Models with cylinder supplies have a cylinder wrench attached to
the system.
•
Models with cylinder supplies have a reserve supply of O2
connected to the machine during system checkout.
•
The necessary emergency equipment is available and in good
condition.
•
Equipment for airway maintenance, manual ventilation, tracheal
intubation, and IV administration is available and in good
condition.
•
Applicable anesthetic and emergency drugs are available.
•
If an auxiliary O2 flowmeter is present, ensure there is adequate
flow.
•
If an optional suction regulator is present, ensure proper
connections and that there is adequate suction.
•
The brakes are set and prevent movement.
•
The power cord is connected to a wall outlet. The mains indicator
comes on when AC power is connected. If the indicator is not on,
the system does not have mains (electrical) power. Use a
different outlet, close the circuit breaker, or replace or connect the
power cable.
•
The O2 flush button is in working condition.
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...