10 Specifications and Theory of Operation
M1132382
10-3
1. Auxiliary O2, 0-10 l/min (optional)
36. Scavenging pressure relief valve, 1.0 kPa
(10 cmH2O)
2. O2 flush
3. 241 kPa (35 psi) second O2 regulator
37. Gas to scavenging: 0-10 l/min drive gas, 0-10 l/min
patient and fresh gas, 0-20 l/min total typical flow
4. Paw Gauge
5. O2 P-Line
38. 200 mL reservoir
6. O2 cylinder (optional)
39. Control bleed approximately 1.0 l/min at 0.29 kPa
(3.0 cmH2O) if continuous (rate dependent)
7. Air P-Line (optional)
8. Total flow sensor (optional)
40. O2 Flush pressure switch
9. Air cylinder (optional)
41. ACGO selector valve
10. N2O P-Line (optional, standard U.S. variant)
42. 22 mm port (ACGO)
11. 758 kPa (110 psi) relief
43. O2 sensor
12. N2O cylinder (optional)
44. Inspiratory flow sensor
13. Cylinder supply and Pipeline supply
45. Absorber
14. Ventilator drive gas select
46. Drain
15. Venturi (optional)
47. Negative pressure relief valve (-14 cmH2O)
16. System switch
48. Sample gas return connection
17. O2 secondary regulator 207 kPa (30 psi)
49. Bag/Vent switch
18. Link-25
50. Bag
19. Flowmeter module (single flow tubes optional)
51. APL valve
20. Gas inlet valve assembly
52. Gas monitor (optional external gas monitor)
21. Air flow meter assembly (optional)
53. Expiratory flow sensor
22. Selectatec manifold
54. Inspiratory flow transducer
23. 37.9 kPa (5.5 psi) pressure relief valve
55. Expiratory flow transducer
24. Vaporizer
56. Sensor interface board
25. N2O balance regulator (required with N2O)
57. Scavenger flow indicator (optional)
26. O2 Flush regulator 172 kPa (25 psi) at
15 l/min
58. Patient
59. 30 mm male - to disposal system
27. Inspiratory flow control valve
60. 0.05 kPa (0.5 cmH2O) entrainment
28. Drive gas check valve (3 cm H2O)
61. Passive gas scavenging interface
29. To disposal system
62. Adjustable scavenging system (optional)
30. Active gas scavenging interface (optional)
63. Filter
31. Mechanical overpressure valve, 10.8 kPa
(110 cmH2O)
64. High or low flow restrictor
65. Bellows
32. Free breathing check valve
66. Scavenging reservoir
33. Room Air
67. Scavenging over pressure relief valve
34. Popoff valve (4 cmH2O)
68. Scavenging reservoir bag with 30 mm male connector
35. Exhalation valve 0.20 kPa (2.0 cmH2O bias)
69. Needle valve assembly (with DISS EVAC connector)
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...