Aespire View
6-10
M1132382
Alarm ranges
Alarm tests
Test the system to verify that alarms are functioning:
1. Connect a test lung to the patient connection.
2. Start a case.
3. Set the Bag/Vent switch to Vent.
4. Set the controls.
•
Ventilation mode: Volume Control (VC).
•
Ventilator.
— Tidal Vol: 400 ml
— Rate: 12
— I:E Ratio: 1:2
— Plimit: 40 cmH2O
— PEEP: Off
•
Anesthesia machine.
— O2 flow: minimum flow (25-75 ml/min).
— All other gases: Off
— Push O2 flush button to fill bellows.
5. Set the O2 concentration to 30% and allow the O2 reading to
stabilize.
Alarm
Range
Increment
Default
O2% high
Off, 21-99%
1%
Off
O2% low
18-99%
1%
21%
MVexp high
Off, 0.5-30 l/min
0.5 l/min
10.0 l/min
MVexp low
Off, 0.1-10 l/min
0.1 l/min
2.0 l/min
TVexp high
20-1600 ml
20 ml
1000 ml
TVexp low
Off, 5-1500 ml
5 ml for less than
20 ml,
otherwise 20 ml
Off
Leak Audio
On, Off
N/A
On