M1132382
Breathing system problems
System
Problem
Solution
Gas scavenging flow is
too low or too high.
Scavenging extract flow
problem.
Use a different scavenging
extraction system. Verify flow is
within specification.
Filter blockage. Active systems
have a flow indicator.
Replace the filter. See
“Remove the AGSS Receiver
Filter”
in the
“ABS Cleaning
and Sterilization”
manual.
The bellows fills when the
Bag/Vent switch is set to
Bag or the bag fills when
the switch is set to Vent.
Leak through the Bag/Vent
switch.
Contact a Datex-Ohmeda
trained service representative
to repair the system.
The ventilator does not
read the position of the
Bag/Vent switch.
Ventilator or absorber
malfunction.
Ventilate manually. Contact a
Datex-Ohmeda trained service
representative to repair the
system.
APL valve does not
operate correctly.
APL valve problem.
Replace APL valve seal and
diaphragm.
Large breathing system
leak not quickly located.
Bag hose not connected
properly.
Ensure that the bag hose is
connected to the bag port
(below the APL valve).
Absorber canister not installed
correctly.
Reinstall the absorber canister,
ensure both pins are engaged.
Bellows falls below top of
indicator during
Leak in the breathing system.
Check, clean or reposition the
pressure relief valve. If the
problem persists, replace the
pressure relief valve, bellows
base, or bellows assembly.
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...