5 Preoperative Tests
M1132382
5-13
Breathing system tests
1. Make sure that the auxiliary equipment is functioning correctly.
2. Verify that AGSS is functioning correctly.
•
Some breathing systems with active AGSS have a flow
indicator on the side. Make sure that the flow indicator shows
a flow in the green range.
3. Make sure that the check valves on the breathing circuit module
work correctly.
•
The expiratory check valve rises during expiration and falls at
the start of inspiration.
•
The inspiratory check valve rises during inspiration and falls
at the start of expiration.
WARNING
Objects in the breathing system can stop gas flow to the
patient, causing injury or death. Use a test plug that is the
appropriate size so that it will not fall into the breathing
system.
After performing the breathing system tests, make sure
that there are no test plugs or other objects caught in the
breathing system.
Bellows test
1. Set the System switch to Standby.
2. Set the Bag/Vent switch to Vent.
3. Set all flow controls to minimum flow.
4. Occlude the patient connection.
5. Push the O2 flush button to fill the bellows. Release the O2 flush
button.
6. Make sure that the pressure does not increase to more than
15 cmH2O on the pressure gauge.
7. If the bellows falls lower than the top of the indicator, there is a
leak. See
in the
section.
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...