10 Specifications and Theory of Operation
M1132382
10-21
SIMV-PC
1. Mandatory pressure control breath
2. Spontaneous pressure supported breath
3. Paw
4. Psupport
5. PEEP
6. Time
Figure 10-7 • SIMV-PC diagram
Synchronized Intermittent Mandatory Ventilation with Pressure
Controlled Ventilation is a mode in which a relatively slow mandatory
breathing rate is set with pressure-controlled breathing. This mode
combines mandatory breaths with spontaneous breath support. If a
trigger event occurs within the synchronization window, a new
pressure-controlled breath is initiated. If a trigger event occurs
elsewhere during the expiratory phase, a support for a spontaneous
breath is provided with pressure support added as set by the clinician.
SIMV-PC settings include:
•
Pinsp - sets the target airway pressure.
•
RR - sets the rate of mechanically driven breaths.
•
Tinsp - sets the inspiration time for a mechanical breath.
•
Psupport - sets the pressure support level.
•
Pmax - sets the maximum airway pressure.
•
PEEP - sets the positive end expiratory pressure.
•
Rise Rate - sets the amount of time to attain Pinsp.
•
End of Breath - the drop in inspiratory flow from the peak
inspiratory flow level where the ventilator stops pressure support
mechanical inspiration and begins exhalation.
•
Trigger - sets the flow trigger level.
•
Trig Window - sets the range in percent of the exhalation phase
where a patient may trigger a mechanical breath.
AB
.9
1
.04
2
T
I
2
1
3
4
5
6
Summary of Contents for Aespire View
Page 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Page 16: ...Aespire View 1 8 M1132382...
Page 46: ...Aespire View 3 16 M1132382...
Page 50: ...Aespire View 4 4 M1132382...
Page 88: ...Aespire View 7 8 M1132382...
Page 112: ...Aespire View 9 10 M1132382...
Page 148: ...Aespire View I 4 M1132382...