5 Preoperative Tests
M1132382
5-15
Monitor and ventilator tests
1. Connect a test lung to the patient connection.
2. Set the Bag/Vent switch to Bag.
3. Set the System switch to On.
4. Push the Menu key.
5. Select
Ventilation Mode
-
VCV
.
6. Set the ventilator parameters.
•
TV 400 ml.
•
Rate 12.
•
I:E 1:2.
•
Pmax 40 cmH2O.
•
PEEP Off.
7. Set the Bag/Vent switch to Vent.
8. Push the O2 flush button to fill the bellows.
9. Make sure that:
•
Mechanical ventilation starts.
•
A subatmospheric pressure alarm does not occur.
•
The ventilator shows the correct data based on settings.
•
The bellows inflate and deflate during mechanical ventilation.
10. Set the O2 flow control to 5 l/min.
11. Make sure that:
•
The end expiratory pressure is approximately 2 cmH2O.
Positive end expiratory pressure that occurs when PEEP is
Off may indicate that the scavenging system is not removing
enough gas.
•
The ventilator shows the correct data based on settings.
•
The bellows inflate and deflate during mechanical ventilation.