WARNING
Datex-Ohmeda strongly recommends the use of O2
monitoring with this equipment. Refer to local standards
for mandatory monitoring.
w
European, international, and national standards require
the following monitoring be used with this system:
•
Exhaled volume monitoring.
•
O2 monitoring.
•
CO2 monitoring.
•
Anesthetic agent monitoring be used when anesthetic
vaporizers are in use.
w
Always make sure that the pipeline supply hoses and the
breathing circuit components are not toxic and will not:
•
Cause an allergic reaction in the patient.
•
React with the anesthetic gases or agent to produce
dangerous by-products.
w
To prevent incorrect values or equipment malfunction,
use only cables, hoses, and tubing approved by
Datex-Ohmeda.
w
This system operates correctly at the electrical
interference levels of IEC 60601-1-2. Higher levels can
cause nuisance alarms that may stop mechanical
ventilation.
w
To help prevent false alarms from devices with high-
intensity electrical fields:
•
Keep the electrosurgical leads away from the
breathing system, flow sensors, and O2 cell.
•
Do not allow the electrosurgical leads to contact any
part of the anesthesia system.
•
Do not use cell phones near the anesthesia system.