O2 cell replacement
WARNING
Handle and dispose of O2 cells according to site
biohazard policies. Do not incinerate.
1. Pull the latch to unlock the flow sensor module.
2. Pull the flow sensor module out of the breathing system.
3. Remove the cable connector from the O2 cell, and unscrew the
cell counterclockwise.
Important
Make sure that the o-ring is on the replacement O2 cell before
installation.
4. Install the replacement O2 cell and reconnect the O2 cell cable.
5. Put the flow sensor module back into the system and push the
latch closed to secure the flow sensor module.
6. Perform “
O2 cell calibration
” after O2 cell replacement.